Select your language

Aerosol Generator ATM 222

Manufacturer: Topas
application sheet PTI/PEMS
product sheet ATM 222

The aerosol generator ATM 222 combines selected technical features of the generators ATM 220 and ATM 228 for long-term stable generation of test aerosols, calibration aerosols or process aerosols from pure liquids (oils like DEHS or PAO, water), solutions (salt solution, sugar solutions) and suspensions (A2, PSL). The generator complies to all requirements of VDI 3941-2.

Standards

VDI 4258-1
VDI 3491-2
DIN EN 61577-4

The generator can be operated via a pneumatic coupling plug by compressed air or other pressurised technical inert-gases (e.g. with N2, CO2). For the purpose of a reproducible and long-term stable aerosol generation, the operational air flow is controlled via the pressure drop over the nozzle. This allows also a reproducible and stable operation of the generator even at very low air flow rates and thus also at lowest particle production rates.

Benefits

  • stable aerosol generation even at low flow rates
  • low noise operation via compressed air or other technical inert-gases (e.g.: nitrogen, carbon dioxide)
  • safe and reproducible adjustment of operating points
  • remote control via serial interface (optional)

Applications

  • aerosol generation from pure liquids, solutions and suspensions
  • fractional efficiency testing of filter media and filter elements
  • generation of long-term stable test and process aerosols
  • calibration and validation of particle measurement techniques

Accessories

  • interface cable
  • aerosol substance (DEHS, PAO, PSL)
  • diffusion dryer (DDU 570/L, DDU 570/H)
Parameter titleUnitValue
setting parameter - differential nozzle pressure
setting range hPa 0 ... 6000
setting resolution hPa 1
aerosol, volumetric flow rate L/h 50 ... 250
aerosol, mass flow rate (aerosol substance) g/h max. 2,0
aerosol, particle production rate #/s 4,0 × 10⁰⁶ ... 1,4 × 10¹⁰
aerosol, particle size range µm 0,01 ... 2
aerosol, distribution width (geometric standard deviation) - 1,8 ... 2,2 (polydispers)
power supply - 12 V DC; 0,5 A
operating medium, gas/air - particle free, dry compressed air or pressurised inert gas
operating medium, gas/air - compressed air supply kPa max. 800
operating medium, aerosol substance - pure liquids (DEHS, PAO, paraffine oil, ...), solutions (NaCl, KCl, ...), suspensions (PSL, A1)
operating medium, aerosol substance - filling volume mL 10 ... 80
environmental condition, max. counterpressure kPa 20
environmental condition, temperature °C 0 ... 50
hose connector, outlet mm Ø 8 (outer diameter)
dimensions (w × h × d) mm 300 × 120 × 195
weight kg 3,0
Copyright MAXXmarketing GmbH
JoomShopping Download & Support