Select your language

SCHMIDT Flow Sensor SS 20.415 LED

Manufacturer: Schmidt
Product brochure (02/2024)
Instructions for use (551742.02D)
3D-STP data
UK-Declaration of conformity (01/2024)

Thermal flow sensor for uni- or bidirectional measurement of air and gases with simultaneous detection of flow direction.

  • Thermopile head (very fast response time)
  • Flow velocity and switching signals
  • Probe length straight version:

- Default: 300 mm

- Special length (option): > 300 … 1,000 mm (step: 10 mm)

  • Probe length of angled version: 150 / 270 mm x 300 mm
  • Measuring ranges 0 … 1 / 2.5 / 10 m/s
  • Bidirectional measurement (optional)
  • Optimized adjustment for LF-applications (optional)
  • Extremely compact (electronics are integrated in probe tube)
  • Sensor element protected against mechanical stress
  • Linear analog output for measured flow velocity
  • Two switching outputs to signal flow direction and threshold
  • Operating status indication and serial number are visible, even when sensor is installed
  • Special version for laminar-flow monitoring in cleanrooms with quick-mounting system for simple installation beneath ceilings or at walls

Option: Monitoring of admissible speed interval (prescribed by official standards) of 0.45 m/s ± 20 % with optical signaling at the sensor (LF status indicator)

  • Laminar flow monitoring in cleanrooms
  • Flow monitoring in pharmaceutical production, semiconductor manufacturing, food production, LCD manufacturing and optical industry
  • Integration into insulators, FFUs and flow boxes

Accessories

Connection technology

DescriptionArticle-Number
Coupler socket, 7-pin with soldering sleeves for cable Ø 0.14 mm² 535 278
Connection cable with coupler socket, 7-pin, 5 m length, open cable ends, material PUR 535 279
Connection cable with coupler socket, 7-pin, length selectable (2 … 100 m; 1 m steps), open cable ends 505 911 - 4
Connection cable with angled socket, 7-pin, 10 m length, open cable ends, material PUR 535 281
Power supply unit: output 24 V / 1A DC, supply voltage 115/230 V AC (strip terminal) 535 282

Mounting and fixing technology

DescriptionArticle-Number
Assembly kit for pipe assembly suitable for MD 10.010/10.015 including pipe clamps and collar for adjustment to pipe diameter 531 394

Measured value display

DescriptionArticle-Number
LED display MD 10.010 in wall housing to show volume flow and flow velocity, 85 … 250 V AC and sensor power supply 527 320
LED display MD 10.010, similar to 527 320 but with 24 V DC voltage supply 528 240
LED display MD 10.015, similar to 527 320 but with additional sum function and second measuring input 527 330
LED display MD 10.015, similar to 527 330 but with 24 V DC voltage supply 528 250

Miscellaneous

DescriptionArticle-Number
Calibrator MK 40.415 - Set (8 / 12 / 16 mA) 546 741 - 4
Technical data
Measuring quantity Standard flow velocity wN, based on standard conditions 1)
Measured medium Clean a orr, nitrogen, additional gases on request
Measuring range wN (±)0 … 1 / 2.5 / 10 m/s (uni- or bidirectional)
Max. display range wN +10 % above measuring range
Lower detection limit wN (±)0.05 m/s
Measuring accuracy
Standard wN ±(3 % of measured value + 0.05 m/s) 2)
High precision (optional) wN ±(1 % of measured value + 0.04 m/s) 2)
Response time (t90) wN 1 s (configurable: 0.01 ... 10 s)
Operating temperature
Sensor 0 … +60 °C
Storage temperature -20 … +85 °C
Material
Sensor head Stainless steel 1.4404
Sensor tube Stainless steel 1.4404
Plug connector, nut screw Stainless steel 1.4404
Light ring PC
General data
Humidity Measuring mode: Non-condensing (< 95 % RH)
Operating pressure (max.) Atmospheric (700 … 1,300 hPa)
Supply voltage 24 V DC ± 10 %
Current consumption Typ. 35 mA / max. 150 mA (including all signal output currents)
Analog output wN 0 … 10 V (RL > 10 kΩ; CL ≤ 10 nF; short-circuit protected)
4 … 20 mA (RL < 300 Ω; CL ≤ 100 nF)
Switching outputs 2 x open collector (at GND), current-limited and short-circuit protected, configurable
-Channel 1 (OC1): Flow direction or threshold value
- Channel 2 (OC2): Threshold value
- External switching voltage / current: Max. 26.4 V DC / 55 mA (typ. 50 mA)
- Hysteresis: 5 % of threshold value (min. ±0.05 m/s)
(<26.4 V DC / 65 mA)
Channel 1 (OC1): direction or threshold value
Channel 2 (OC2): threshold value
Switching hysteresis 5 % of the threshold value, min ± 0.05 m/s
Error signal Only current interface: 2 mA (according to NAMUR NE43)
Electrical connection Plug-in-connector M9, male, 7-pin (shielded), screwed
Maximum cable length Voltage interface: 10 m; current interface: 100 m
Mounting position Straight: ±3° relative to flow direction (uni- or bidirectional)
Angled: Vertical downdraft flow (per design
Protection type IP65; electrical connector: IP67 (with correctly attached cable)
Protection class III (SELV or PELV)
Sensor length Angled: 150/270 mm (vertical) x 300 mm (horizontal)
Straight: Standard 300 mm; special length >300 … 1,000 mm
Weight Approx. 200 g (angled design with 270 mm x 300 mm)