- Sanitize the R2S(Sampler) and place at desired test location.Ensure that vacuum and power cable connections are secure
The R2S Microbial Air Sampler incorporates industry standard Slit-to-Agar methodology with comparable microbial recovery/sensitivity, offering a highly accurate analysis of air quality. The R2S is designed for monitoring clean rooms, LAF Hoods and Isolators, but can be employed for monitoring in virtually any critical environment. It is designed to allow for sanitization of all surfaces exposed to the environment while its streamlined design affords minimal disruptive effects on Laminar Airflow in critical environments.
- Constructed of Aluminum / 316 Stainless Steel
- Exterior Finish of White Epoxy/Polyester Coating
- Clear Polycarbonate Dome Assembly w/Distance Gauge
- Weight of 3.25 Pounds
- Dimensions of 5.5″ Height x 5″ Diameter
- Turntable is operated by a Stepper Motor (V100)
- Base is Chemically Sanitizable, Dome/Seal Autoclavable
- From 0-120 Minutes of Active Monitoring per Sample Cycle
- Operates at Sample Rates of 28.3, 50 & 100 LPM
- Works with the V100 Controller
- Generates virtually no particulate matter
- Uses standard 90mm Test Plates
- Allows for the estimated time of organismal recovery

- Remove the dome assembly, retaining it over the sampler to minimize the risk of accidental contamination (false positive)

- Place the 100mm Test Plate (TSA) on turntable and remove the test plate lid

- Replace the dome assembly and ensure that dome is properly seated in the seal. Place the test plate lid next to sampler on a sanitizedurface

- Use the turntable adjustment knob to set the proper distance between the test plate and sample throat (2-3mm) as achieved when just touching the recieved indicator

- Ensure that correct sampling device is selected, and set desired test volume and test period,and start the sample cycle.

- Perform sampling during designated critical processing step or Quality control test being performed

- When test period is complete stop sample cycle, or the sample cycle will self terminate when desired test volume is obtained

- Remove the dome assembly and replace the test plate lid

- Remove the test place from the sampler, secure with tape and label with appropriate sample parameters.

- Replace the dome assembly to keep sample chamber contaminant free until the next test plate is employed.Place generated sample run on the test plate and submit to lab for procesiing