Select your language

SCHMIDT Flow Sensor SS 23.400 ATEX

Manufacturer: Schmidt
Product brochure (02/2023)
Instructions for Use (514622.02H)
Type examination certificate (valid as of 09/2013)
3D-STP data
EU-Declaration of conformity (01/2024)
UK-Declaration of conformity (01/2024)
Type examination certificate (valid until 09/2013)

Thermal flow sensor for uni- or bidirectional measurement of flow rate of air and gases with simultaneous detection of the flow direction in an explosive atmosphere (Zone 2).

  • Thermopile head (very fast response time)
  • Flow velocity and switching signals
  • Probe length 130 … 300 mm
  • Measuring ranges 0 … 1 m/s to 0 … 20 m/s
  • Bidirectional measurement (optional)
  • Detection of flow direction (optional)
  • Sensor programming (optional)
  • Cleanroom sensor
  • Extremely compact (electronics is integrated in sensor tube)
  • Sensor element protected against mechanical stress
  • Linear analog output for measured flow velocity
  • Two switching outputs to signal flow direction and threshold
  • ATEX gases, zone 2: II 3G Ex nA IIC T4 Gc
  • Laminar flow monitoring in cleanrooms
  • Monitoring of overflow
  • Cooling air monitoring
  • Flow measurement in test benches
  • Mass and volume flow measurement

Accessories

Connection technology

DescriptionArticle-Number
Connection cable with coupler socket, 5 m length, open cable ends, material PUR 535 279
Power supply unit: output 24 V / 1A DC, supply voltage 115/230 V AC (strip terminal) 535 282

Mounting and fixing technology

DescriptionArticle-Number
Wall mounting flange, stainless steel 1.4571 520 181
Compression fitting stainless steel, G½, atmospheric pressure 532 160
Wall mounting bracket, aluminium, anodized 503 895
Mounting flange, stainless steel 1.4404 551 740
Assembly kit for pipe assembly suitable for MD 10.010/10.015 including pipe clamps and collar for adjustment to pipe diameter 531 394

Measured value display

DescriptionArticle-Number
LED display MD 10.010 in wall housing to show volume flow and flow velocity, 85 … 250 V AC and sensor power supply 527 320
LED display MD 10.010, similar to 527 320 but with 24 V DC voltage supply 528 240
LED display MD 10.015, similar to 527 320 but with an additional sum function and a second measuring input 527 330
LED display MD 10.015, similar to 527 330 but with 24 V DC voltage supply 528 250

Miscellaneous

DescriptionArticle-Number
Calibrator MK 40.415 - Set (8 / 12 / 16 mA) 546 741 - 4
Technical data
Measuring quantity Standard flow velocity wN based on standard conditions 1)
Measured medium Clean air or nitrogen, additional gases on request
Measuring range wN (±)0 … 1 / 2.5 / 5 / 10 / 20 m/s (uni- or bidirectional)
Lower detection limit wN (±)0.05 m/s
Measuring accuracy
Standard wN ±(3 % of measured value + [2 % of fmr; min. 0.05 m/s]) 2)
High precision (optional) wN ±(1 % of measured value + [2 % of fmr; min. 0.04 m/s]) 2)
Response time (t90) wN 1 s (configurable: 0.01 … 10 s)
Operating temperature
Sensor 0 … +60 °C
Storage temperature -20 … +85 °C
Material
Sensor head Stainless steel 1.4404
Sensor tube Stainless steel 1.4404
Plug-in connector, nut screw Stainless steel 1.4404
General data
Humidity Measuring mode: Non-condensing (< 95 % RH)
Operating pressure (max.) Atmospheric (700 … 1,300 hPa)
Supply voltage 12 … 26.4 V DC
Current consumption Typ. < 35 mA (max. 150 mA; including signal current of switching outputs)
Analog output wN 0 … 10 V (RL > 10 kΩ; CL ≤ 10 nF; short-circuit protected)
4 … 20 mA (RL < 300 Ω; CL ≤ 100 nF)
Switching outputs 2 x open collector (at GND), current-limited and short-circuit protected, configurable
-Channel 1 (OC1): Flow direction or threshold value
- Channel 2 (OC2): Threshold value
- External switching voltage / current: Max. 26.4 V DC / 55 mA (typ. 50 mA)
- Hysteresis: 5 % of threshold value (min. ±0.05 m/s)
Electrical connection Plug-in-connector M9, male, 7-pin (shielded), screwed
Maximum cable length Voltage interface: 15 m;
Current interface: 100 m
Mounting position Arbitrary
Protection type Sensor probe: IP66
Plug-in connection: IP67 (with correct attached cable)
Protection class III / PELV (EN 50178)
ATEX declaration Gas (zone 2): II 3G Ex nA IIC T4 Gc
Probe length 130/200/300 mm
Weight Approx. 60 g max. (300 mm Sensor length)
Copyright MAXXmarketing GmbH
JoomShopping Download & Support